Milwaukee PH56 Waterproof pH/ºC/ºF Tester
User Manual
Milwaukee Instruments Martini PH56 Waterproof pH//ºC/ºF Tester, Meter, PH 56
pH DUAL LEVEL LCD TESTER WITH AUTOMATIC TEMPERATURE COMPENSATION, 2 POINT AUTOMATIC CALIBRATION, � 0.01 pH ACCURACY; RANGE: -2.00 to 16.00 pH.
pH/temperature
Specifications:
Range -2.00 to 16.00 pH
-5.0 to 60.0 �C
Resolution 0.01 pH
0.1 �C
Accuracy @ 20�C � 0.05 pH
�0.3 �C
Calibration automatic, 1 or 2 points
temperature compensation automatic
Probe (included) pH/�C (Mi56P)
Auto off after 8 minutes of non-use
Dimensions 200 x � 38mm
Weight 100 g
Accessories
Milwaukee Instruments MI56P pH Temp Probe
Replacement spare pH probe electrode sensor for PH55 PH56 pH digital pen tester meter, MI 56P
Martini MI56 56 $36.94 | Suggested Solutions for pH Meters / Controllers
1. MA9015 storage solution for pH probe storage.
2. MA9016 cleaning solution for pH probe cleaning.
3. MA9007 pH 7 buffer for 1-point pH calibration.
4. MA9004 pH 4 and MA9010 pH 10 for 2-point pH calibration.
After each use, always rinse the probe with water, clean it with MA9016 cleaning solution and store the probe with MA9015 storage solution.
For probe cleaning - Soak it in MA9016 cleaning solution for about 3 minutes once a day or at least once a week to prevent clogging and to maintain accuracy.
For probe storage - Place a sponge or tissue in the bottom of the cap, and pour some MA9015 storage solution into the cap.
Never let probe dry after conditioning and Never use distilled/deionized water for storage purposes.
Always rinse the pH electrode with water between calibrations and testings.
Check out the "pH Electrode Care" for more details. |